USFDA WHO GMP Guidelines Pharmaceutical Regulations Information about Pharma Quality Managment System Validations Process Validation Cleaning Validation Equipment Qualification Equipment Process Validation Computer System Validation Sterile Product Manufacturing Guidelines Media fill guideline Aseptic Techniques in Sterile dosage form Clean Room Clean Room Classification Guidelines on Area Qualification HVAC AHU HEPA Filter HEPA Filter integrity testing and leak test

Tuesday, August 11, 2009

ENVIRONMENTAL MICROBIAL MONITORING OF ASEPTIC AREA AIR SAMPLING METHOD

OBJECTIVE :- The purpose of this procedure is to describe the method for sampling a measured quantity of air so that viable particles ( cfu ) are impinged upon an agar surface
to allow subsequent culture and quantitification



UNIT OF RESPONSIBILITY : QA Executive and trwained QA Assistants

MATERIAL :-

1 Air Sampler

2 Sterile soyabean casien digest agar strips


Note :- The media preparation batch no is given to prepared media



IMPORTANT POINTS/NOTE/PRECAUTIONS
:-

1 It is not advisable to expose the Agar strips beyond 7 minutes as the medium gets
desiccated

2 The coated surface of the Agar strips should face down during incubation or while
storing in order to avoid water of condensation


3 The instrument can be operated either on battery (6 dry cell batteries of 105 Volts)
or with battery eliminator of 9 volts & 1500 m amp

4 Ensure that the air flow rate is calibrated and certified



PROCEDURE:-

1 Prepare the Soyabean casien digest agar media as per media preparation SOP

2 Autoclave the media at 121 deg C for 20 minutes

3 After completion of autoclaving transfer the media to the microbiology room

4 Prepare the strips by pouring the sterilized Soyabean casien digest agar medium

aseptically taking care to avoid contamination from the person preparing them. Cool the strips

5 The prepared strips are labelled and incubated for 24 hours at 35 to 37 degree C

6 Sterilize the drum, impeller blade, timer box, cylinder the use of 70 % IPA and fumigation

respectively

7 Tear open the plastic tubing covering Agar strip

8 Slide back the sliding cover filted on the outer socket Remove the agar strip with the

medium filled side facing downwards & insert into the slot in the side of the drum of

the Air Sampler

9 Care must be taken that the medium does not touch the impeller and the end strip

produces enough for it to be pulled out of the instrument again after the sampling



10. To avoid contamination do not touch the layer of culture medium in the agar strips

before and after measuring process

11 Bring the Air Sampler to the required position Set up the running time and

switch on


12 The sampling time may be set as per the following table

Switch

1 ON 1 Min = 40 lts

2 ON 2 Min = 80 lts

4 ON 4 Min = 160 lts



All Switches On 7 minutes = 280 lts

13 Remove and incubate the agar strip for 72 hours at 35 to 37 deg C

15 After incubation count the cfu on the agar strip , and identify the growth observed on the

agar strip their gram character by Gram staining refer sop



14 For immediate second sampling using same drum & impeller blade, 70 % IPA may be

used for sterilization

15 Record the observation in the attachment 1



OTHER RELEVANT DETAILS:-

a) EVALUATIONS :-

Total no of colonies observed on the strip be counted with the help of colony counter or

directly by holding the strip against light The total microbial load may be arrived by using

the following formula -
Total no. of colonies on agar strip
c.f.u./litre of air = -----------------------------------------------------------
Volume of air sucked in X time of exposure
Total no. of colonies
c.f.u/ m /min = ------------------------------------------ X 25
Time of exposure in minutes
Limits
1) Manufacturing area : : NMT 10 cfu
2) Storage Area : : NMT 10 cfu
3) Filling area : : NMT 05 cfu
Standard operating procedure SOP for microbiology and aseptic techniques for manufacturing of sterile dosage forms injections

This Website is a Guide for Pharmaceutical Manufacturing Pharmacy Students Pharmacy Colleges and Pharmacists pharmaceutical companies in health care
Following are some of articles which will be useful for you in further understanding of aspects of sterile dosage form manufacturing

Pharmaceutical Validation

Cleen Room Classification

Pharmaceutical Industry

Pharmaceutical Manufacturing

Clinical Trials

Media fill

Pharmaceutical companies

Regulatory Affairs in Pharmaceutical Industry

No comments:

Pharma Guidelines Most Read

.

Readers Visit Number

This website is Authentic Source of Pharmaceutical Guidelines

This website is Authentic Source of Pharmaceutical Guidelines

Pharma Validations Quality Management System USFDA Guidelines WHO GMP Guidelines New Drug Update





This website consist of detail Information about Pharma Guidelines Pharma Regulatory Affairs USFDA Guidelines WHO Guidelines Pharmaceutical Quality Management System Pharma Validations Process Validation Cleaning Validation Equipment Qualification Equipment Process Validation Computer System Validation Sterile Product Manufacturing Guidelines Media fill guideline Aseptic Techniques in Sterile dosage form Clean Room Clean Room Classification Guidelines on Area Qualification HVAC AHU HEPA Filter HEPA Filter integrity testing and leak test